Measurement uncertainty in calibration depends not only on the calculation method but also on the completeness of the measurement model and the relative contribution of individual sources across the measuring range. This study investigates uncertainty sources in the calibration of a capacitive relative humidity hygrometer by comparison with a reference chilled-mirror hygrometer in a climatic chamber at 25 °C. An eight-input measurement model was evaluated at 20, 50, and 80 %RH using the GUM uncertainty framework, with results validated by the Monte Carlo method. The uncertainty budget changed substantially with humidity. At 20 %RH, calibration and drift of the reference instrument accounted for 72% of the combined variance. At 80 %RH, hysteresis of the instrument under calibration became the largest contribution at 45%, followed by reference calibration and chamber non-uniformity. Expanded uncertainty increased from 0.70 to 1.37 %RH across the range. Monte Carlo validation confirmed the GUM interval at 20 %RH, was borderline at 50 %RH, and failed at 80 %RH because the dominant rectangular hysteresis component produced a non-Gaussian output distribution. Targeted reduction of hysteresis and chamber non-uniformity decreased expanded uncertainty by up to 30%. The findings support point-specific uncertainty budgets and Monte Carlo validation when non-Gaussian components dominate.
Keywords
measurement uncertainty, calibration, uncertainty budget, relative humidity, Monte Carlo method, GUM.
References
1. Castro, H. F. F. (2024). Mathematical modeling applied to the uncertainty analysis of a tank prover calibration: Understanding the influence of calibration conditions on the GUM validation using the Monte Carlo method. Flow Measurement and Instrumentation, 96, Article 102547. https://doi.org/10.1016/j.flowmeasinst.2024.102547
2. García-Berrocal, A., Montalvo, C., Carmona, P., & García-Álvarez, R. (2024). Optimization of the Monte Carlo simulation to improve the calibration uncertainty of volume meters for hydrocarbons custody transfer. Applied Sciences, 14(24), Article 11472. https://doi.org/10.3390/app142411472
3. Joint Committee for Guides in Metrology. (2023). Guide to the expression of uncertainty in measurement—Part 1: Introduction (JCGM GUM-1:2023). BIPM. https://doi.org/10.59161/JCGMGUM-1-2023
4. Kapić, A., Tsirou, A., Verdini, P. G., & Carrara, S. (2023). Uncertainty analysis of polymer-based capacitive relative humidity sensor at negative temperatures and low humidity levels. Measurement, 209, Article 112468. https://doi.org/10.1016/j.measurement.2023.112468
5. Possolo, A., Hibbert, D. B., Stohner, J., Bodnar, O., & Meija, J. (2024). A brief guide to measurement uncertainty (IUPAC Technical Report). Pure and Applied Chemistry, 96(1), 113–134. https://doi.org/10.1515/pac-2022-1203
6. Schaude, J., & Hausotte, T. (2024). Uncertainty-based determination of recalibration dates. International Journal of Metrology and Quality Engineering, 15, Article 2. https://doi.org/10.1051/ijmqe/2023017
7. van der Veen, A. M. H., Cox, M. G., & Possolo, A. (2022). GUM guidance on developing and using measurement models. Accreditation and Quality Assurance, 27, 295–297. https://doi.org/10.1007/s00769-022-01509-8
How to Cite
Kazakov Tolibjon (2026). INVESTIGATION OF THE SOURCES OF MEASUREMENT UNCERTAINTY IN THE CALIBRATION OF MEASURING INSTRUMENTS. Science and Innovation, 5(9), 86-90. https://doi.org/10.67620/sai.59a0012